UPC/EAN: 860012796212
Meets EN/IEC 61000-4-4 & ISO7637
ECL50 is designed for pulse calibration with a 50Ω load at the front input and a matching 50Ω output impedance which meets and exceeds EN/IEC 61000-4-4 & ISO7637. While the specification exceeds 400MHz (875ps rise time), our design at the 3dB bandwidth is over 500MHz achieving 700ps to ensure you have no doubt of capturing your pulse rise time (10% - 90%).
Specification
Max. power rating: 5 Wrms
Max. peak voltage (burst/EFT pulses as specified in EN/IEC 61000-4-4): 10.0 kV
Max. operating time at 10.0 kV: 300 s (5 kHz, 300ms/15ms)
Nominal attenuation (terminated by 50 Ω): -60dB
Nominal voltage ratio (See exact values on labels): 1000:1
Input Impedance: 50 Ω ±2%
Output Impedance: 50 Ω ±1%
Bandwidth (3 dB): >400 MHz
Connector Input: SHV connector, female
Connector Output: BNC jack, female