UPC/EAN: 860012796212
 

  • Meets EN/IEC 61000-4-4 & ISO7637

 

ECL50 is designed for pulse calibration with a 50Ω load at the front input and a matching 50Ω output impedance which meets and exceeds EN/IEC 61000-4-4 & ISO7637. While the specification exceeds 400MHz (875ps rise time), our design at the 3dB bandwidth is over 500MHz achieving 700ps to ensure you have no doubt of capturing your pulse rise time (10% - 90%). 

 

Specification 

Max. power rating: 5 Wrms

Max. peak voltage (burst/EFT pulses as specified in EN/IEC 61000-4-4): 10.0 kV

Max. operating time at 10.0 kV: 300 s (5 kHz, 300ms/15ms)

Nominal attenuation (terminated by 50 Ω):  -60dB

Nominal voltage ratio (See exact values on labels): 1000:1

Input Impedance: 50 Ω ±2%

Output Impedance: 50 Ω ±1%

Bandwidth (3 dB): >400 MHz

Connector Input: SHV connector, female

Connector Output: BNC jack, female

ECL50 | 50Ω EFT/Burst Attenuator : EN/IEC 61000-4-4 & ISO7637

SKU: ECL50 - RENT
$2,000.00 Regular Price
$300.00Sale Price
Only 5 left in stock
Price Options
Weekly Rental
$100.00every week until canceled
30 Day Rental
$200.00every month until canceled
    bottom of page